SME SYSTem
1038.6002.02 3.119 E-13
:SYSTem:KLOCk ON | OFF
The command (Keyboard LOC
k) disables the keyboard of the SME including the [LOCAL] key or
enables it again (OFF).
Caution: If :SYSTem:SECurity is ON, the keyboard cannot be enabled, i.e., :SYSTem:KLOCk
OFF is not accepted. If the disabling of the command is released by switching over
to :SYSTem:SECurity OFF, data will be lost.
Example: :SYST:KLOC ON *RST value is OFF
:SYSTem:MODE FIXed | MSEQence
The command sets the operating mode of the instrument.
FIXed The overall instrument state can only be switched over using *RCL.
MSEQuence The instrument successively sets the instrument states indicated under
:SYSTem:MSEQuence:RCL. *RST value is FIXed
Example: :SYST:MODE FIX
:SYSTem:MSEQuence
This node follows the SOURce:LIST system. It can manage several memory sequences which
each consist of a list of instrument state numbers and a time list. If :SYSTem:MODE is switched
to MSEQuence, the instrument states stated in the list selected are set successively for the time
stated in the time list in each case.
:SYSTem:MSEQuence:CATalog?
The command queries the memory sequences available. It returns a list, the entries are
separated by means of commas.
Example: :SYST:MSEQ:CAT? Response: "SEQ1", "DEMO", "SEQA"
:SYSTem:MSEQuence:DELete "Name of sequence"
The command deletes the memory sequence indicated.
Example: :SYST:MSEQ:DEL "SEQ1"
:SYSTem:MSEQuence:DELete:ALL
The command deletes all memory sequences. The memory-sequence mode must be switched
off as a selected sequence cannot be deleted (SYSTem:MODE FIXed).
Example: :SYST:MSEQ:DEL:ALL
:SYSTem:MSEQuence:DWELl 50 ms to 60 s{,50 ms to 60 s}
For the memory sequence which has currently been selected, the command transmits a list
indicating the time for which an instrument setting is "held" in each case before the instrument
proceeds to the next setting. If DWELl indicates only one parameter, every item of the instrument
state list is set for the same, indicated time. Lists are not influenced by *RST.
Example: :SYST:MSEQ:DWEL 1s