EasyManua.ls Logo

Rosemount 2088 - International; B.1.5 International

Rosemount 2088
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Certifications
March 2017
Reference Manual
00809-0100-4108, Rev CA
Product Certifications
94
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm.
Installation, maintenance and use shall take into
account the environmental conditions to which the
diaphragm will be subjected. The manufacturer’s
instructions for installation and maintenance shall be
followed in detail to assure safety during its expected
lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from
electrostatic discharge. Avoid installations that could
cause electrostatic build-up on painted surfaces, and
only clean the painted surfaces with a damp cloth. If
paint is ordered through a special option code,
contact the manufacturer for more information.
I1 ATEX Intrinsic Safety
Certificate: BAS00ATEX1166X
Standards: EN60079-0:2012 + A11:2013,
EN60079-11:2012
Markings: II 1 G Ex ia IIC T4 Ga (–55 °C
T
a
+70 °C)
Special Condition for Safe Use (X):
1. The apparatus is not capable of withstanding the 500
V insulation test required by EN60079-11. This must
be taken into account when installing the apparatus.
N1 ATEX Type n
Certificate: BAS00ATEX3167X
Standards: EN60079-0:2012 + A11:2013,
EN60079-15:2010
Markings: II 3 G Ex nA IIC T5 Gc (–40 °C
T
a
+70
°C)
Special Condition for Safe Use (X):
1. The apparatus is not capable of withstanding the 500
V insulation test required by EN60079-15. This must
be taken into account when installing the apparatus.
ND ATEX Dust
Certificate: BAS01ATEX1427X
Standards: EN60079-0:2012 + A11:2013,
EN60079-31:2009
Markings: II 1 D Ex t IIIC T50 °C T
500
60 °C Da
Special Conditions for Safe Use (X):
1. The user must ensure the maximum rated voltage and
current (36 volts, 24 milliamps, d.c.) are not
exceeded. All connection to other apparatus or
associated apparatus shall have control over this
voltage and current to a category 'ib' circuit.
2. Cable entries must be used which maintain the
ingress protection of the enclosure to at least IP66.
3. Unused cable entries must be filled with suitable
blanking plugs which maintain the ingress protection
of the enclosure to at least IP66.
4. Cable entries and blanking plugs must be suitable for
the ambient range of the apparatus and capable of
withstanding a 7 J impact test.
5. The Rosemount 2088/2090 sensor module must be
securely screwed in place to maintain the ingress
protection of the enclosure.
B.1.5 International
E7 IECEx Flameproof
Certificate: IECEx KEM 06.0021X
Standards: IEC 60079-0:2011, IEC60079-1:2014,
IEC60079-26:2014
Markings: Ex d IIC T6...T4 Ga/Gb, T6(–60 °C
T
a
+70
°C), T4/T5 (–60 °C T
a
+80 °C)
Special Condition for Safe Use (x):
1. This device contains a thin wall diaphragm.
Installation, maintenance and use shall take into
account the environmental conditions to which the
diaphragm will be subjected. The manufacturer’s
instructions for installation and maintenance shall be
followed in detail to assure safety during its expected
lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from
electrostatic discharge. Avoid installations that could
cause electrostatic build-up on painted surfaces, and
only clean the painted surfaces with a damp cloth. If
paint is ordered through a special option code,
contact the manufacturer for more information.
I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 12.0071X
Standards: IEC60079-0:2011, IEC60079-11:2011
Markings: Ex ia IIC T4 Ga (–55 °C
T
a
+70 °C)
Table 1. Input Parameters
Parameter
HART
®
Voltage U
i
30 V
Current I
i
200 mA
Power P
i
0.9 W
Capacitance C
i
0.012 μF

Table of Contents