Specifications and Reference Data
77
Specifications and Reference Data
March 2017
Reference Manual
00809-0100-4108, Rev CA
Storage
Humidity limits
0-100% relative humidity.
Volumetric displacement
Less than 0.0005 in
3
(0,008 cm
3
).
Damping
Analog output response time to a step change is
user-selectable from 0 to 60 seconds for one time constant.
Software damping is in addition to sensor module response
time.
A.3.8 Physical specifications
Material selection
Emerson
provides a variety of Rosemount product with
various product options and configurations including
materials of construction that can be expected to perform
well in a wide range of applications. The Rosemount
product information presented is intended as a guide for
the purchaser to make an appropriate selection for the
application. It is the purchaser's sole responsibility to make a
careful analysis of all process parameters (such as all
chemical components, temperature, pressure, flow rate,
abrasives, contaminants, etc.), when specifying product,
materials, options and components for the particular
application. Emerson is not in a position to evaluate or
guarantee the compatibility of the process fluid or other
process parameters with the product, options,
configuration or materials of construction selected.
A.3.9 Process connections
Rosemount 2088
1
/2-14 NPT female
DIN 16288 G
1
/2 male
RC
1
/2 female (PT
1
/2 female)
M20 3 1.5 (CM20) male
Rosemount 2090P
1
1
/2 –11.5 NPT or 1-in. flush mount
Rosemount 2090F
1
1
/2-inch or 2-in. Tri Clamp Connection
A.3.10 Process wetted parts
Isolating Diaphragm
Rosemount 2088
316L stainless steel (UNS S31603)
Alloy C-276 (UNS N10276)
Rosemount 2090P
316L stainless steel.
Sensor fill 2088
(1)
1. If storage temperature is above 85 °C, perform a sensor trim prior to
installation
2090F 2090P
Silicone -50 to 230 °F
(-46 to 110 °C)
N/A -50 to 185 °F
(-45 to 85 °C).
For process
temperatures
above 185 °F
(85 °C) below
Inert -50 to 185 °F
(-46 to 85 °C)
N/A N/A
Neobee N/A -22 to 185 °F
(-30 to 85 °C)
N/A
With LCD
display
-40 to 175 °F
(-40 to 80 °C)
(2)
2. For the output code N, LCD display may not be readable and LCD display
updates will be slower at temperatures below -22 °F (-30 °C).
-22 to 175 °F
(-30 to 80 °C)
(3)
3. LCD display may not be readable and LCD updates will be slower at
temperatures below -22 °F (-30 °C).
-40 to 175°F
(-40 to 80
°C)(3)