Page #35 background image
Cubis
®
MCA Operating Instructions 35
Installation
5 Installation
5.1 Scope of Delivery
Item Quantity
Device 1
Weighing pan 1
Shield disk 1
Pan support 1
Power supply unit 1
Country-specific power supply cable with test seal 1
USB connection cable 1
In-use dust cover for control unit 1
In-use dust cover for the weighing module 1
5.2 Selecting an Installation Site
Procedure
t Make sure that the following conditions are met at the installation site:
Condition Features
Ambient conditions Suitability tested (see Chapter “15.2 Ambient
Conditions”, page 74)
Setup surface Stable, even surface that is not exposed to
vibrations
Not directly against a wall
Sufficiently dimensioned for the device and the
peripheral devices (device space requirements see
Chapter “15.1 Dimensions and Weight”, page 74;
peripheral device space requirements see
instructions for the peripheral devices, e.g. printer)
Sufficient load-bearing capacity for the device and
the peripheral devices even when full (device
weight see Chapter “15.1 Dimensions and Weight”,
page 74; weight of the peripheral devices see
instructions for the peripheral devices, e.g. printer)
Access Barrier-free

Table of Contents