MCA Operating Instructions
System Settings
7.5.12 Parameters in the “Connections / SBI Protocol” Menu
Parameter Settings Explanation
Format Value without header The data output only exports the measured value without ID code.
Value (with header)* The data output exports the measured value with ID codes.
Date & time, value (two
lines)
The data output exports the measured value, date, and time.
Output Without stability* The data is output without the weight value being stable.
With stability The data is output only when the weight value is stable.
Automatic data
output
Off* Deactivates the automatic data output. The data output must be triggered
by means of an SBI command with “ESC P”.
Each weight value Starts the automatic data output with each value according to the display
result.
1 second Starts the automatic data output every second.
2 seconds Starts the automatic data output every 2 seconds.
5 seconds Starts the automatic data output every 5 seconds.
10 seconds Starts the automatic data output every 10 seconds.
30 seconds Starts the automatic data output every 30 seconds.
60 seconds Starts the automatic data output every 60 seconds.
* Factory setting
7.5.13 Parameters in the “Connections / Climate Module” Menu
Parameter Settings Explanation
Climate modules
7.5.14 Parameters in the “Connections / Connected Devices / Motion Sensor” Menu (Only for Connected Motion Sensors)
Parameter Settings Explanation
Number of gestures 2 gestures (left, right)* Activates the control of functions with 2 gestures. Gestures: left, right.
4 gestures (left, right, up,
down)
Activates the control of functions with 4 gestures. Gestures: left, right,
down, up.
Any gesture Activates the control of functions with any gesture.
Gesture to the left
Gesture to the right
Gesture upwards
Gesture downwards
Start/cancel taring The gesture starts a taring process or cancels a taring process.
Start/cancel zeroing The gesture starts the zeroing process or cancels the zeroing process.
* Factory setting