66 Cubis
®
MCA Operating Instructions
Operation
8.13.2 Running the “Statistics” Application
The “Statistics” application saves up to 100 weight values and evaluates these
statistically.
The following values are saved and exported by the statistics application:
Number of components
Mean value
Standard deviation
Variation coefficient
Sum of all values
Lowest value (minimum)
Highest value (maximum)
Spread: Difference between maximum and minimum
The “Statistics” application can be combined with the following functions:
Unit Change, only available in the weighing display, before saving the initial
weight value
ID marking
Automatic taring
Procedure
t Open the main menu.
t Start the task for the “Statistics” application.
t Zero the device.
t Place the sample on the weighing pan.
t To start the recording of the statistics: Press the [Confirm] button on the keypad.
y The current weight value is saved and the device is automatically tared.
t To save the next value: Place a new sample on the weighing pan and press the
[Confirm] button.
t To display an overview of the saved data: Press the [Report] button.
t To print the current statistics: Press the [Print] button.
t To exit the current statistics, and to delete the saved values: Press the [Exit]
button.