8.9 Adjusting with the isoCAL Function
The device can be automatically internally calibrated and adjusted using the isoCAL
function.
Requirements
The device is not located in the menu.
Alphanumeric inputs are not active.
The load on the weighing pan remains unchanged for 2 minutes.
The load on the weighing pan amounts to no more than 2% of the maximum
load.
The device does not register an input for 2 minutes.
If all requirements for starting the isoCAL function are met and one of the following
conditions occurs: The isoCAL function is automatically triggered.
Possible conditions are:
The ambient temperature has changed since the last adjustment/calibration.
The interval time was exceeded (interval time, see Chapter “15.9 isoCAL
Function”, page 80).
The device has been leveled.
The device has been disconnected from the power supply since the last
adjustment/calibration (only for conformity-assessed models).
Procedure
t If the isoCAL automatic start function is set and the [isoCAL] button is displayed
as the predominant button in the operating display:
t Wait until the isoCAL function starts automatically.
y In the operating display, a time display counts down to 0.
y If no load change or no operation takes place on the device before the
expiration of the time display: The isoCAL function starts.
t If the isoCAL manual start function is set and the [isoCAL] button is displayed as
the predominant button in the operating display:
t Press the [isoCAL] button.
t If information is displayed stating that the weighing pan is loaded: Unload
the weighing pan.
y The isoCAL function starts.
y If the isoCAL function is complete: The device confirms the completion of the
calibration/adjustment process with an acoustic signal, and the calibration report
is displayed.
t To close the calibration report and return to the previous display: Press the [OK]
button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA10202S Series and is the answer not in the manual?

Sartorius Cubis MCA10202S Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA10202S Series
CategoryScales
LanguageEnglish

Related product manuals