60 Cubis
®
 MCA Operating Instructions
 Operation
 t If an individual application from the displayed QAPP package is to be activated:
 t Press the desired application.
 y A display opens containing details about the selected application.
 t Press the [License] button.
 y The input field for the license key appears.
 t If an additional cost is associated with the application: Enter the license key 
in the entry field and press the [OK] button.
 t If no additional cost is associated with the application: Press the [OK] button.
8.4.2  Adding an Application to a Task
Applications must be added to a task so that they can run.
Procedure
 t Open Task Management.
 t Press the [New] button.
 y A list of all activated applications is displayed.
 t To select an application: Press the desired application.
 y The wizard for creating a new task starts.
 t Follow the wizard’s instructions in the operating display.
8.5  Adding Weighing and Print Profiles to a Task
To be able to use a weighing or print profile: Add a weighing or print profile to a 
task. Weighing and print profiles can be configured in the Settings menu.
Procedure
 t Open the task.
 t Create or edit a task.
8.6  Preparing Weighings
The device must be prepared before every weighing.
Procedure
 t Level the device.
 t Zero the device. In order to do so, press the [Zero] button.
 t If the device cannot be zeroed: Remove the sample to be weighed and re-zero 
the device.
 t Calibrate the device.