15.8 Recommended Calibration Weight
MCA14202S MCA14202P MCA10202S MCA8202S
Unit Value Value Value Value
External test weight g 14000 14000 10000 7000
Recommended accuracy class E2 E2 E2 E2
MCA6202S MCA6202P MCA4202S MCA2202S
Unit Value Value Value Value
External test weight g 5000 5000 3000 1500
Recommended accuracy class E2 E2 E2 E2
MCA1202S MCA12201S MCA8201S MCA5201S
Unit Value Value Value Value
External test weight g 700 12000 8000 5000
Recommended accuracy class E2 F1 F1 F1
15.9 isoCAL Function
15.9.1 Models MCA14202S | MCA14202P | MCA10202S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling
15.9.2 Models MCA8202S | MCA6202S | MCA6202P | MCA4202S | MCA2202S | MCA1202S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 2
After a time interval h 12
After successful leveling
15.9.3 Models MCA12201S | MCA8201S | MCA5201S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 4
After a time interval h 12
After successful leveling