Cubis
®
 MCA Operating Instructions  55
    System Settings 
7.5.15  Parameters in the “Device Settings / Date and Time” Menu
Parameter Settings Explanation
NTP configuration NTP operating status Activates or deactivates the time synchronization with the NTP server.
Server IP Saves the entered server ID for the NTP server.
Set the date and time Date format Sets the date display format to DD/MM/YYYY, MM/DD/YYYY, DD.MM.YYYY, 
YYYY-MM-DD (ISO)*, YYYY:MM:DD, DD-MMM-YYYY or MMM-DD-YYYY.
Set date Saves the entered date.
Time format Sets the time display format to HH.MM.SS, HH:MM:SS (ISO)* or HH:MM:SS 
am/pm.
Set time Saves the entered time.
Time zone Saves the selected time zone.
* Factory setting
7.5.16  Parameters in the “Device Settings / Safe Weighing” Menu
Parameter Settings Explanation
isoCAL execution 
mode
Off Deactivates the isoCAL function. This setting change is not available for all 
models.
Show status field, manual 
start
If the device must be calibrated: The [isoCAL] button is displayed as the 
predominant button in the operating display. The isoCAL function must be 
manually triggered using the [isoCAL] button.
On, automatic execution* Activates the isoCAL function. The device is automatically adjusted as soon 
as a trigger starts the isoCAL function.
On, automatic execution 
with linearization (only if 
possible)
Activates the isoCAL and linearization function. The device is automatically 
adjusted and then linearized as soon as a trigger starts the isoCAL function.
Safety level Off If the isoCAL function must be carried out: The device displays a status 
message.
Warning* If the isoCAL function must be carried out: The device displays a warning 
message. Some device functions are limited until isoCAL has been 
executed: The displayed weight value is marked as invalid; printed data are 
marked with [!].
Strict If the isoCAL function must be carried out: The device displays an error 
message. Some device functions are limited until isoCAL has been 
executed: The displayed weight value is marked as invalid; the starting of 
functions and saving of data in tasks is disabled; data output is 
deactivated.
* Factory setting