Operation
8.10 Internally Calibrating and Adjusting the Device
Requirements
The weighing pan is unloaded.
Procedure
t Open the main menu.
t Press the “Internal Adjustment” task.
y The internal calibration / adjustment function is executed.
y If automatic leveling is set for a model with motorized adjusting feet: The device
levels itself automatically.
y If the calibration / adjustment function is complete: The device confirms the
completion of the calibration / adjustment process with an acoustic signal, and
the calibration report is displayed.
t To close the calibration report and return to the main menu: Press the [OK]
button.
8.11 Weighing and Printing with ID Marking
8.11.1 Saving Values for Printouts
Every lot and every sample can be assigned an ID number. The ID numbers are saved
in the print memory and exported during the printing process.
Requirements
The prompt for the lot IDs and sample IDs is activated for the “Weighing” task.
Procedure
t Open the main menu.
t Start the “Weighing” task: To do so, tap on the task.
y The input field for the lot ID is displayed.
t If a barcode scanner is connected to the device: Enter the lot ID in the entry field
or scan using the barcode scanner.
t If no barcode scanner is connected to the device: Type the lot ID into the entry
field.
t Press the [OK] button.
y The weighing display is displayed.
t Zero the device. In order to do so, press the [Zero] button.
t Place the sample on the weighing pan.
t Press the [Confirm] button.
y The input field for the sample ID is displayed.
t Type the sample ID into the entry field.
t Press the [OK] button.
y The [Print memory] button appears in the operating display.
y The weight value and entered IDs are saved.
t If additional values are to be saved:
t Remove the sample being weighed.
t Place the next sample on the weighing pan and press the [Confirm] button.
t Type the sample ID into the entry field.
t Press the [OK] button.