Cubis
®
 MCA Operating Instructions  77
    Technical Data 
15.7  Metrological Data
15.7.1  Models  MCA14202S | MCA14202P | MCA10202S | MCA8202S
MCA14202S  MCA14202P MCA10202S MCA8202S
Unit Value Value Value Value
Scale interval (d)
mg 10 10 | 20 | 50 10
10
Maximum capacity (Max) g 14200 3500 | 7000 |  
14200
10200 8200
Repeatability at 5% load
Standard deviation of the load values, 
tolerance
mg 10 10 7 7
Standard deviation of the load values, 
typical value
mg 5 5 5 4
Repeatability at approx. maximum capacity
Standard deviation of the load values, 
tolerance
mg 10 10 7 7
Standard deviation of the load values, 
typical value
mg 5 5 5 4
Linearity deviation
Tolerance mg 30 50 20 20
Typical value mg 10 20 6 6
Deviation when load is off-center,  
positions according to OIML R76
Test weight g 5000 5000 5000 5000
Tolerance mg 20 40 20 30
Typical value mg 10 10 10 10
Sensitivity drift between +10°C and +30°C ppm/K 1.5 1.5 1.5 2
Tare maximum capacity: Less than 100 % of 
maximum capacity
Accuracy class according to Directive 2014/31/EU I I I II
Verification scale interval (e) according to 
Directive 2014/31/EU
g 0.1 0.1 0.1 0.1
Minimum load (Min) according to Directive 
2014/31/EU
g 1 1 1 0.5
Minimum initial weighing according to USP 
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 8.2 8.2 8.2 8.2
Typical minimum initial weighing g 8.2 8.2 8.2 8.2
Typical stabilization time s 0.8 0.8 0.8 1
Typical measurement time s 1.5 1.5 1.5 1.5