Cubis
®
MCA Operating Instructions 79
Technical Data
15.7.3 Models MCA1202S | MCA12201S | MCA8201S | MCA5201S
MCA1202S MCA12201S MCA8201S MCA5201S
Unit Value Value Value Value
Scale interval (d)
mg 10 100 100
100
Maximum capacity (Max) g 1200 12200 8200 5200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 7 50 50 50
Standard deviation of the load values,
typical value
mg 4 20 20 20
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 7 50 50 50
Standard deviation of the load values,
typical value
mg 4 20 20 20
Linearity deviation
Tolerance mg 20 100 100 100
Typical value mg 6 30 30 20
Deviation when load is off-center,
positions according to OIML R76
Test weight g 500 5000 5000 2000
Tolerance mg 20 200 200 200
Typical value mg 10 100 100 100
Sensitivity drift between +10°C and +30°C ppm/K 2 4 4 4
Tare maximum capacity: Less than 100% of
maximum capacity
Accuracy class according to Directive 2014/31/EU II II II II
Verification scale interval (e) according to
Directive 2014/31/EU
g 0.1 1 1 1
Minimum load (Min) according to Directive
2014/31/EU
g 0.5 5 5 5
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 8.2 82 82 82
Typical minimum initial weighing g 8.2 82 82 82
Typical stabilization time s 0.8 0.8 0.8 0.8
Typical measurement time s 1 1 1 1