7.5.4 Parameters in the “Weighing and Print Profiles / Weighing” Menu
Parameter Settings Explanation
Ambient conditions Very stable Sets the ambient conditions to “very stable”: Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
Stable Sets the ambient conditions to “stable”.
Recommended for the following work environment:
Stable table
Slight movement in the room
Slight draft
Unstable* Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate.
Recommended for the following work environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a significantly
delayed change in the weight values and long wait for stability with a
further reduction in the output rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
Application filter Weighing (final readout)* Activates a filter that enables a fast change in the display for very fast load
changes.
Display changes with minimal load changes (in the digit range) occur more
slowly.
Dosing (filling mode) Activates a filter that enables a very fast change in the display with
minimal load changes (e.g. when filling containers).
Sensor mode (low filtering) Activates a weak but fast filter that always behaves in the same way for
load changes (e.g. when filling automated systems).
Dynamic mode (without
filtering)
Deactivates the active application filter.
Stability Very high accuracy Sets the stability to “very high accuracy”.
High accuracy Sets the stability to “high accuracy”.
Medium accuracy* Sets the stability to “medium accuracy”.
Fast Sets the stability to “fast”.
Very fast Sets the stability to “very fast”.
Very slow Sets the stability to “very slow”.
* Factory setting for the “Standard Weighing” factory-set weighing profile

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA10202S Series and is the answer not in the manual?

Sartorius Cubis MCA10202S Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA10202S Series
CategoryScales
LanguageEnglish

Related product manuals