Chapter 8 Service Window
GC-2014 Service Manual
87
8.8 Fan & LCD Use Time
8.9 Init. for Shipment
This window is used to reset the operation times of the
column-oven fan and the LCD unit.
Note: The settings for the [Acceptable value] cannot be
changed.
Fan: 61320.0 hours (approx. 7 years)
LCD: 46380.0 hours (approx. 5 years)
Operation beyond the period indicated by these acceptable
values is possible, but a result of "Not Good" will appear in
self-diagnostic reports.
After replacing the column-oven fan or the LCD unit, reset
the corresponding operation time.
pf2: Pressing [Reset] clears the counter at the cursor position.
This window is used only when being shipped from the
factory.
CAUTION
Do not use this window.

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)

Related product manuals