12 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH
Important Information
Subject to change without notice
1.2.4 Hazards due to electrostatic discharges
The electronic housing of the FLSE100 and of the optionally available meter body is
painted by the manufacturer with a layer thickness of max. 0.2 mm.
WARNING:
Ignition hazards through electrostatic discharges exist when the FLOWSIC100
Flare with special paint and a layer thickness >
0.2 mm is used in applications
with ignition group IIC in accordance with ATEX and IECEx.
For installation, the risk of electrostatic charging of the surface must be
reduced to a minimum.
Use appropriate caution when performing maintenance and cleaning work.
For example, the surfaces should only be cleaned with a damp cloth. The
affected devices will be identified by the manufacturer by a warning label.