168 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH
Start-up and Parameter Settings
Subject to change without notice
4.2.3 Configuration of application-specific parameters for 10“ pipelines and 2-path
configurations
In order to set application-specific parameters for 10“ pipelines and 2-path configurations,
log into SOPAS via access level 2 - "Service".
Establish connection between MCUP and SOPAS ET using USB interface.
Log in as "Service" (password "service").
Fig. 103 Service Login
Set SOPAS to maintenance mode.
Fig. 104 Maintenance mode page
WARNING:
No other parameters or settings may be changed while logged into access level
2 - "Service", except for those described in this section. All other changes to
settings may only be carried out using the access levels 0 - "Operator" or 1 -
"Authorized client". The manufacturer is not liable for any disruption caused by
incorrect settings made by the user in SOPAS.