Start-up and Parameter Settings
FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH 195
Subject to change without notice
4.3.5.2 Setting Ethernet module parameters
Assigning the Ethernet module a new IP address
An IP address specified by the customer is entered at the factory if the address is available
when the device is ordered. If not, standard address 192.168.0.10 is entered.
The following steps are necessary to change the address:
Set the desired network configuration in the "Parameter / System Configuration" direc-
tory, Field "Interface module Ethernet" field.
Click "Reset module" in the "Interface Module - Type Selection".
Fig. 131 "Parameter / System Configuration" directory
Assign the new IP address to SOPAS ET (
pg. 155, § 4.1.3).
New address
NOTICE:
Malfunctions in data transfers not caused by the measuring system can
occur during communication via Ethernet.
The FLOWSIC100 Flare manufacturer assumes no responsibility for
malfunctions that may occur during equipment operation when measured
value transfers and their usage to control processes run solely via
Ethernet.
Increasing the value in the "Scantimeout" field to 3000 ms can minimize
communication problems.