Product Description
FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH 67
Subject to change without notice
Fig. 27 Connection cables
Specific requirements for selection and use of cables (esp. in hazardous areas)
pg. 119, § 3.8.1.
NOTICE:
Only analog cable Exi of device type FL100 EX-S is approved for ATEX and
CSA versions (interconnection of master and slave unit).
All other cables offered by SICK are only for ATEX/IECEx versions for use in
ATEX zone 1 or zone 2 applications.
All cable glands offered are only for ATEX/IECEx versions for use in ATEX
zone 1 or zone 2 applications.
Cables, cable glands and other installation material for CSA Cl I, Div 1 ver-
sions must be provided by the customer. Solutions for CSA Cl I, Div2/zone 2
versions are available on request.