68 CLINITEK Status+ Analyzer Operator’s Guide
Troubleshooting
E11 Failure of test table The test table is positioned
improperly.
1. Make sure that the test table is
in place.
2. Move the test table in or out of
the analyzer slightly to
reposition the test table.
3. If the error remains, with the
analyzer powered on,
disconnect the power cord from
the back of the analyzer and
connect it back in. Press the on/
off button to power on the
analyzer.
4. If the error remains with the test
table in place, contact your local
technical support provider or
distributor. See Appendix B,
Support Information, When to
Contact Technical Support
page 131.
E12 Failure of LED An LED light source failed.
Contact your local technical support
provider or distributor.
E20 Failure of clock Contact your local technical support
provider or distributor.
Error
Code
Error Message Action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens CLINITEK Status+ and is the answer not in the manual?

Siemens CLINITEK Status+ Specifications

General IconGeneral
DisplayLCD
Sample TypeUrine
Measurement MethodReflectance photometry
TypeUrinalysis Analyzer
Test TypesUrine Chemistry
ConnectivityRS232 port
Weight5.0 kg
Power Supply100-240 V AC, 50-60 Hz
Data StorageUp to 200 patient results
Test ParametersLeukocytes, Nitrite, Urobilinogen, Protein, pH, Blood, Specific Gravity, Ketone, Bilirubin, Glucose
Test StripsMultistix and other compatible strips