CLINITEK Status+ Analyzer Operator’s Guide 71
Troubleshooting
E53 Strip Test selected
but cassette
detected
Repeat the test using the Cassette
Test routine (see Section 2,
Operations, Performing an hCG
Cassette Quick Test‚ page 35).
E54 Cassette Test
selected but strip
detected
Repeat the test using the Strip Test
routine (see Section 2, Operations,
Performing a Urinalysis Strip Quick
Test‚ page 29).
E56 Incorrect size test
table
Repeat the test using the correct test
table (see Section 2, Operations,
Performing a Urinalysis Strip Quick
Test‚ page 29).
E57 Missing strip or
cassette
Repeat the test and ensure that you
correctly position the strip or
cassette on the test table (see
Section 2, Operations, Performing a
Urinalysis Strip Quick Test‚ page 29
or Performing an hCG Cassette Quick
Test‚ page 35).
E58 Misplaced strip 1. Repeat the test and ensure that
you correctly position the strip
on the test table (
see Section 2,
Operations, Performing a
Urinalysis Strip Quick Test‚
page 29).
2. If the error remains, examine the
test table insert to ensure that
the small, white line located
near the tip of the strip (on the
strip side of the insert) is present
and not damaged.
3. If this line is damaged, contact
your local technical support
provider or distributor.
Error
Code
Error Message Action