72 CLINITEK Status+ Analyzer Operator’s Guide
Troubleshooting
E59 Inverted strip
positioned on the
test table
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2, Operations, Preparing a
Urinalysis Strip Quick Test‚ page 29).
E60 Tilted strip Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2, Operations, Preparing a
Urinalysis Strip Quick Test‚ page 29).
E61 Dry strip Repeat the test with a fresh strip and
ensure that the strip has been in
contact with the sample (see Section
2, Operations, Preparing a Urinalysis
Strip Quick Test‚ page 29).
E62 Light Ingress Too much light is reflecting on the
analyzer. Move the analyzer to a
location with lower lighting.
Contact your local technical support
provider or distributor.
E63 Failure to find end
of strip
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2, Operations, Preparing a
Urinalysis Strip Quick Test‚ page 29).