Service software 3 - 11
Siemens AG SPL1-130.840.02 Page 11 of 18 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Explanation:
The displayed position (distance and position) of the objects on the ultrasound image is
calculated from the time difference between sending a signal and receiving the echo.
The standard speed of sound in the human body is 1540 m/sec, however, this speed can
vary depending on which tissue is being examined.
The position of the iso-center can be adjusted to ensure that the position of the cross mat-
ches the actual stone position on the ultrasound monitor.
This is possible with:
Fig. 1 Fig. 2
The following values can only be altered if approval is given by
Erlangen.
The correct values can be found in document SPL1-130.038.01..
(Intranet).
Convex
Array
Sector
CAUTION
change values for sector applicator - Press key 's'
change values for convex applicator - Press key 'c'
SECTOR APPLICATOR
-----------------------------------------
actual values change with key (-) (+)
-----------------------------------------
threshold value = 100 mm t T
scale low = 7.7 % l L
scale high = 7.7 % h H