LITHOSTAR MODULARIS 4 - 29
Siemens AG SPL1-130.840.02 Page 29 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Perform the following check:
- position the support arm horizontally and place it in the docking position (center locking
position).
- place a level on the support arm (1/Fig. 35).
- the air bubble can point to the right, but not to the left (Fig. 35). If the air bubble moves
to the left side, the spring must be replaced.
Check the iso-center with radiation.