LITHOSTAR MODULARIS 4 - 29
Siemens AG SPL1-130.840.02 Page 29 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
• Perform the following check:
- position the support arm horizontally and place it in the docking position (center locking
position).
- place a level on the support arm (1/Fig. 35).
- the air bubble can point to the right, but not to the left (Fig. 35). If the air bubble moves
to the left side, the spring must be replaced.
• Check the iso-center with radiation.