12 Software description
CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032 141
To stop the acquisition process for that sample and save the acquired
data, click the Stop button. Acquisition will move on to the next defined
Worklist item to be acquired. If the current sample is the last in the
acquisition sequence, acquisition will terminate.
Restart: The button is only enabled if acquisition is in progress. Click the
button to restart the acquisition process for that sample. The data
acquired up to this point are discarded.
Stop: The button is only enabled if acquisition is in progress. Click the
button to stop the acquisition process for that sample. The data acquired
up to this point are saved and acquisition will move on to the next defined
Worklist item to be acquired. If the current sample is the last in the
acquisition sequence, acquisition will terminate.
Previous: The button is disabled if acquisition is in progress. Click the
button to activate the previous Worklist item in the Worklist.
Next: The button is disabled if acquisition is in progress. Click the button
to activate the next Worklist item in the Worklist.
Abort: The button is only enabled if acquisition is in progress. Click the
button to stop acquisition and discard data from the current sample. You
will have the option to clear the acquired data and proceed to the next
Worklist item or to clear the acquired data and stop measuring.
The Acquire pane contains the following buttons for performing predefined
procedures:
Prime: Click the button to start a priming procedure. Once started, the
software guides you through the process.
De-bubble: Click the button to start a procedure to clear bubbles from
the flow cuvette. Once started, the software guides you through the
process.
Clean: Click the button to start a cleaning procedure. Once started, the
software guides you through the process.
The "Flow Rates" section of the Acquire pane contains the following controls:
Above the flow rate slider, the acquisition flow rate (events per second)
and the absolute event count are displayed during acquisition.
Flow Rate Slider: The slider allows to adjust the sample injection speed
into the flow cuvette. Low speed values result in a better precision and
accuracy. A higher speed can be used when the particle concentration is
measured and accuracy is of lower relevance. The scale ranges from
0.1 μl/s to 19.8 μl/s. The value (µl/s) is displayed at the right-hand side of
the slider.
The flow rate can be changed in the following ways:
Mouse drag and drop of the slider will increase/decrease the flow rate in
steps of 0.1 μl/s.