6 Operation
36 CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032
Defining the Worklist for remaining samples
Defining a Worklist for future measurements includes defining the
Workspace, instrument settings and stop conditions. Instrument settings and
stop conditions are loaded from the Workspace by default but can be
changed in the Worklist.
Procedure
1. In the Worklist, enter a Lab # and sample IDs for the Worklist items as
desired.
2. To apply the Workspace developed from the sample to all Worklist items,
right-click the Worklist item to open the context menu and select
"Workspace > Apply To".
– A dialogue appears.
3. Select whether to apply the Workspace to the current panel or to the
entire Worklist.
4. If differing instrument settings are to be used, click into the "Instrument
Settings" cell of the Worklist item, select "Load from file" and select the
desired instrument settings file.
5. To apply the differing instrument settings to other Worklist items,
right-click the "Instrument Settings" cell to open the context menu and
select "Apply to Panel".
6. To define differing stop conditions for the remaining samples, select the
desired stop conditions from the Stops pane (see chapter 12.5.6 Stops
pane on page 144).
Or right-click the "Stop Conditions" and "Stop Value" columns in the
Worklist to open the context menu and select the desired stop conditions.
– The Worklist for remaining samples is defined.
Saving the Worklist
The FCS data are saved automatically on completion of sample
measurement. The Worklist can be saved separately.
Procedure
1. To save the Worklist, select the File tab of the Ribbon (see chapter
12.4.1 File tab on page 58) and select "File > Save > Save Worklist" or
"Save Worklist As".
– The Worklist is saved.