11 Technical data
CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032 47
Operating
environment
Temperature:
Operation: 15 °C to 30 °C
Transport and storage: 5 °C to 50 °C
Relative humidity:
Operation: 20 % to 85 %, non-condensing
Transport and storage: 20 % to 85 %,
non-condensing
Room:
Clean environment. Direct sun light should be
avoided
Operation specifications
Setup time Max. 5 minutes
Parameters detected Up to 8 optical parameters: FSC, SSC, FL1, FL2,
FL3, FL4, FL5, FL6
Particle size range 0.1 μm to 100 μm
Maximum data 15000 events/second
Data acquisition Max. 15000 events/second
Acquisition stop
condition
Event-based or volume-based
Trigger On all parameters, on multiple parameters or on
single trigger parameter, selectable in software
Data resolution 16 bit
Application specifications
Application Immunophenotyping, DNA analysis, apoptosis,
microbiology, industrial applications, 3 to 6 colour
analysis.
True Volumetric Absolute Counts = counting per
volume
True Volumetric
Absolute Counting
Based on precise counting and mechanical fluid
volume measurement no need for reference sample
or beads.
Hardware data