Performance  Check  Procedure
DISPLAY
Equipment  Required  (See  Table 7-1)
Time  mark generator  50 Cl  BNC  coaxial  cable
50 Cl  BNC termination
1.  TRACE  ROTATION
a.  Set:
READOUT  (Intensity) 
A  INTEN
VERTICAL  MODE 
CH  1  VOLTS/DIV 
CH  1  COUPLING 
SCOPE  BW 
A/B  SELECT 
TRIGGER MODE 
TRIGGER SOURCE 
TRIGGER CPLG 
TRIGGER SLOPE 
TRIGGER HOLDOFF 
TRIGGER LEVEL 
Horizontal  MODE 
Horizontal  POSITION 
A SEC/DIV 
CLEAR MEAS’ MT
FOCUS
For a viewable  readout
For a viewable trace
CH  1
0.1  V
AC
Off
A Trigger 
AUTO  LEVEL 
VERT 
DC
_j~  (positive-going)
Min
12  o'clock 
A
12  o'clock 
2  |xs
Press to  remove  measurement 
cursors.
For best  defined  display
b.  Position trace vertically to the  center graticule  line.
c.  CHECK—trace  rotation  control  range  is  adequate  to  align  trace  with 
center graticule  line  using  a small  straight-bladed  alignment tool.
d.  ADJUST—trace  parallel to  center horizontal  graticule  line.
7-8
2245A  Operators