Performance  Check  Procedure
17.  CH  3  and  CH  4  Vertical  Bandwidth
a.  Set:
VERTICAL  MODE  CH  3  (CH  2  off)
CH  3  and CH  4 VOLTS/DIV  0.1V
b.  Connect leveled  sine-wave generator  (SG  503)  output to the  CH  3 in
put via a 50 n   precision  BNC coaxial cable  and a 50 f l   BNC termina
tion.
c.  Set the  generator output  for a six-division  signal display  at  50  kHz.
d.  Set  the  generator  Frequency  Range  and  Frequency  Variable  controls 
for a 90  MHz  output  frequency.
e.  CHECK—that  the  signal  display  amplitude  is  4.2  divisions  or  more  as 
the frequency is  increased to  100  MHz.
f.  Repeat the  procedure  for 0.5 VOLTS/DIV  setting.
g.  Move the test  signal to the  CH  4 input.
h.  Set VERTICAL  MODE to  CH 4
i.  Repeat the  procedure  for CH 4.
18.  SCOPE BW  (Bandwidth  Limit)  Accuracy
a.  Set:
VERTICAL  MODE  CH  1  (CH 4  off)
CH  1  VOLTS/DIV  10  mV
SCOPE  BW  On
b.  Move test  signal from  the  CH 4 input to the  CH  1  input.
c.  Set leveled sine-wave generator  (SG 503)  output for a six-division sig
nal amplitude  at  50 kHz.
d.  Increase  the leveled  sine-wave  generator  output  frequency,  using the 
Frequency Range and Frequency Variable controls until a signal display 
amplitude  of 4.2  divisions is  obtained.