Operator Checks  and Adjustments
2.  Set:
VERTICAL  MODE 
Horizontal  MODE
3.  Press the AUTO  SETUP
As desired  (see NOTE) 
As desired 
button.
NOTE
Normally,  the  Auto  Setup  of  the  display  Is  sufficient  for 
measurement purposes.  If further waveform  enhancement 
is  needed,  use  the  appropriate  front-panel  controls  to 
adjust the display.  For auto-setup action,  see Appendix A.
Trace  Rotation  Adjustment
1.  Perform  the  “ Initial  Setup”  procedure.  Position  the  trace  vertically  to 
align it with the center horizontal graticule line and  check that the trace 
is  parallel with the  graticule line.
NOTE
Normally,  the  trace  will be parallel  to  the  center horizontal 
graticule  line,  and  the  TRACE  ROTATION  adjustment  will 
not be needed.
2.  If necessary,  adjust the TRACE  ROTATION to make the baseline trace 
parallel to the center horizontal graticule line.  Use a small straight-blade 
screwdriver or alignment tool.
Probe  Low-Frequency Compensation
Misadjustment of probe compensation is a possible source of measurement 
error.  The attenuator probes are equipped with compensation adjustments. 
To  ensure the best  measurement  accuracy,  always check probe  compen
sation before  making  measurements.
1.  Connect the two  supplied  10X probes to the  CH  1  and  CH 2  BNC input 
connectors.
2245A  Operators
4-3