Performance  Check  Procedure
INTRODUCTION
This  Performance Check Procedure verifies the Performance Requirements 
of the  2245A as  listed  in  the  Performance  Characteristics  (Section  6)  and 
helps  determine  the  need  for  readjustment.  These  checks  may  also  be 
used  as  an  acceptance  test  or as  a preliminary  troubleshooting  aid.
You  do not  have to remove the wrap-around cabinet from the 2245A to do 
this  procedure.  All  checks  can  be made  with  controls  and  connectors  ac
cessible from the  outside.
Test  Equipment  Required
Table 7-1  lists all the test equipment required to do the Performance Check 
Procedure.  Test equipment specifications described are the minimum nec
essary  to provide  accurate  results.  For test  equipment  operating  informa
tion,  refer to  the  appropriate test  equipment instruction  manual.
When you use  equipment  other than  that  recommended,  you  may  have to 
make some changes  to the test setups.  If the  exact  example equipment in 
Table 7-1  is  not  available,  use  the  Minimum  Specification column to deter
mine  if  any  other  available  test  equipment  might  be  adequate  to  do  the 
check.
Performance  Check  Interval
To ensure instrument  accuracy,  check the performance  of the 2245A after 
every 2000 hours of operation,  or once each year if used infrequently.  If the 
checks indicate a need for readjustment or repair,  refer the instrument to a 
qualified service  person.
Preparation
This  procedure  is  divided  into  subsections  to  let  you  check  individual  sec
tions  of  the  instrument  when  it  is  not  necessary  to  do  the  complete  Per
formance  Check.  An  Equipment  Required  block  at  the  beginning  of  each 
subsection  lists  the  equipment  from  Table  7-1  that  is  needed  to  do  the 
checks  in  that  subsection.
2245A  Operators
7-1