Performance  Check  Procedure
h.  Same as b.  above.
i.  Repeat CHECK for  CH 4.
j.  Set  CH  3  and CH 4 VOLTS/DIV to  0.5 V.
k.  Set calibration generator to  Std Ampl  output,  2  V.
l.  CHECK—the  signal  amplitude  is  between 3.92  and 4.08  divisions.
m.  Set:
VERTICAL  MODE  CH  3  (CH  4  off)
CH  3 VOLTS/DIV  0.5 V
n.  Move the test  signal to the  CH  3  input.
o.  Repeat CHECK  procedure for CH  3.
p.  Disconnect  the  test  setup from the  2245A.
11.  ADD  Mode and  CH 2  INVERT  Deflection  Accuracy
a.  Set:
VERTICAL  MODE  ADD  (all others  off)
CH  1  and  CH  2  VOLTS/DIV  0.1  V
CH  1  and CH  2  Input  COUPLING  DC
b.  Connect  calibration generator  Std Ampl  output to the  CH  1  and  CH  2 
inputs  via  a  50  A   precision  BNC  coaxial  cable  and  a  BNC  dual-input 
coupler.
c.  Set  the  calibration  generator to  Std Ampl  output,  0.2  V.
d.  Position the ADD signal to the center of the  crt graticule with the  CH  1 
and CH  2  POSITION  controls.
e.  CHECK—That the ADD signal amplitude is between  3.92 and 4.08 divi
sions.
f.  Set  CH  2  INVERT On.
g.  CHECK—That  the ADD  signal  amplitude  is  0.08  division  (less than half 
a minor graticule division)  or less excluding trace width  (sweep will free 
run).
h.  Disconnect the  test  setup from  the  2245A.
2245A  Operators
7-17