EasyManua.ls Logo

Tektronix 2245A - 500 Khz Trigger Sensitivity; Khz Trigger Sensitivity

Tektronix 2245A
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Check Procedure
NOTE
The TRIGGER LEVEL control may be used to obtain a stable dis
play.
e. CHECK—that the display Is stably triggered with DC, HF REJ, and AC
Trigger CPLG; and that the display will not trigger on NOISE REJ or LF
REJ Trigger CPLG.
f. Set:
Horizontal MODE
TRIGGER CPLG
A/B SELECT
B TRIGGER MODE
B TRIGGER SOURCE
B TRIGGER SLOPE
B SEC/DIV
DELAY Time
B INTEN
B
DC
B Trigger
NORM
VERT
_ y (positive-going)
0.5 ms
?0.000 (minimum delay time)
For viewable display
NOTE
It may be necessary to adjust the TRIGGER LEVEL control to ob
tain a display.
g. CHECK—that, using the Trigger LEVEL control the display can be
stably triggered in DC, HF REJ, and AC Trigger CPLG; and that the
display cannot be triggered in NOISE REJ or LF REJ Trigger CPLG.
h. Disconnect the test setup from the CH 1 Input.
2. 500 kHz Trigger Sensitivity
a. Set:
SCOPE BW
Horizontal MODE
A/B SELECT
A SEC/DIV
Off
A
A Trigger
2 p.S
b. Connect leveled sine-wave generator (SG 503) output to the CH 1 In
put via a 50 XI BNC coaxial cable and a 50 XI BNC termination.
2245A Operators
7-29

Table of Contents

Related product manuals