Operators  Familiarization
Auto  Setup
Pressing  the  AUTO  SETUP  button  automatically  sets  up  the  front-panel 
controls based on the characteristics of the applied signal.  The voltage am
plitude,  sweep  settings,  trigger  parameters,  vertical  and  horizontal  posi
tioning,  and trace intensities  are preset to  produce  a usable waveform dis
play.  The waveform  is  horizontally  centered  and vertically  positioned within 
the  crt  display.  The  voltage  and  frequency characteristics  of the  input sig
nal must  be within the  limits  of the  2245A specifications  given in  Section  6.
Auto-setup action  for each  front-panel  control  is  shown  in  Appendix A.
MIN  SETUP
MIN  SETUP  is  only  available  on  instruments  Serial  Num
bered B020100 or above.
Pressing and holding the  Auto-Setup button until the instrument  returns to 
normal  operation  initiates  the  MIN  SETUP function.  This automatically sets 
up the front-panel controls but leaves many of the controls unchanged from 
what  the  user has  set.
MIN  SETUP  action for  each front-panel  control  is  whown in  Appendix A.
MEASUREMENT SYSTEM
The 2245A has a highly accurate voltage- and time- measurement system. 
Voltage  or time is  measured  between  two  cursors  (or intensified zones for 
time  measurements  in  the  ALT  or  B  Horizontal  MODE)  that  can  be  posi
tioned,  as  desired,  to two  points  on  the waveform.
Cursor Volts  Measurements
Press  CURSOR  VOLTS  to  display  the  voltage  measurement  cursors  and 
activate  the  voltage  measurement  system.  The  measured  value  is  dis
played  at the  top  of the  screen.
NOTE
Changing  the  Horizontal  MODE  will  remove  the  volts  cur
sors  from  the  display.  Set  Horizontal  MODE  back  to  A  or 
X-Y and press  the  CURSOR  VOLTS  button  to  redisplay the 
cursors.
2245A  Operators
3-5