Performance  Check  Procedure
11.  X -Y   Phase  Differen ce
a.  Set:
HORIZONTAL  MODE 
VERTICAL MODE 
CH  1  Input COUPLING
A
CH  1  (CH  2  off) 
DC
b.  Connect  leveled  sine-wave  generator  output  to  the  CH  1  input  via  a 
50 f l  BNC coaxial  cable  and a 50 f l   BNC termination.
c.  Set leveled sine-wave generator output for six divisions of signal display 
amplitude  at 50  kHz.
e.  Position dot to  graticule center.
f.  Set CH  1  Input COUPLING to  DC.
g.  CHECK—ellipse opening at the center is 0.3 division or less,  measured 
horizontally.
12.  X-Axis  Bandwidth
a.  Set VERTICAL MODE to CH 2  (CH  1  off).
b.  Set  leveled  sine-wave  generator  output  for  six  divisions  of  horizontal 
display amplitude  at  50 kHz.
c.  Set  leveled sine-wave output to 3  MHz.
d.  CHECK—X-Axis  display  is  4.2  horizontal divisions or more.
e.  Disconnect the test  equipment  from the  2245A.
d.  Set:
Horizontal  MODE 
CH  1  Input  COUPLING
X-Y
GND
2245A  Operators
7-45