Performance  Check  Procedure
2.  Geometry
a.  Connect time mark generator  (TG 501)  to CH  1  via a 50 Cl  BNC coax
ial  cable  and a  50 f l  BNC termination.
b.  Set generator for 0.2  p.s  time  marks.
c.  Position the bottom of the CH  1  signal below the  bottom graticule line. 
It may be necessary to increase the A intensify in order to see the time 
markers.
d.  CHECK—deviation of any vertical line within the center eight horizontal 
divisions  does not  exceed  0.1  division  (half a minor division).
e.  Set CH  1  COUPLING to GND.
f.  Position trace slowly from the bottom graticule line to the top graticule 
line while making the following check.
g.  CHECK—bowing  or  tilt  of  baseline  trace  doesn't  exceed  0.1  division 
(half a minor  division)  within the  eight vertical  divisions.
h.  Disconnect  test  signal from the  2245A.
2245A  Operators
7-9