Performance  Check  Procedure
c.  Set:
CH  1  VOLTS/DIV  10 mV
CH  1  VAR VOLTS/DIV  Fully  CCW
d.  Connect  calibration  generator  (PG  506)  Std Ampl  output  to  the  CH  1 
input via a 50 f l   precision  BNC coaxial cable.  Set generator  Std Ampl 
output to 50 mV.
e.  CHECK—the  signal amplitude  is  two  divisions  or less.
f.  Set:
CH  1  VAR VOLTS/DIV  Detent  (calibrated)
Vertical  MODE  CH  2  (CH  1  off)
CH  2  VOLTS/DIV  10 mV
g.  Position CH  2  trace  to  the  center horizontal graticule  line.
h.  Move the test  signal to the CH  2  input.
i.  Set CH 2 VAR VOLTS/DIV fully  CCW.
j.  Repeat the CHECK  procedure for CH  2.
k.  Set CH  2  VAR VOLTS/DIV to  detent  (calibrated)  position.
8.  Low-Frequency  Linearity
a.  Set:
VERTICAL  MODE  CH  1  (CH  2 off)
CH  1  VOLTS/DIV  10  mV
SCOPE  BW  ON
b.  Set  calibration  generator to  Std Ampl  output, 20  mV.
c.  Move the test  signal to  the  CH  1  input.
d.  Position the top  of the  signal to top  graticule  line.
e.  CHECK—the  signal amplitude  is between  1.9 and  2.1  divisions.
f.  Set bottom  of the  signal to  bottom graticule  line.
7-14
2245A  Operators