Performance  Check  Procedure
c.  Set leveled sine-wave generator output to produce a 7.0 division sine- 
wave  display amplitude  at 500 kHz.
d.  Add a 10X and a 2X BNC attenuator before the  50 f l  BNC termination 
(for a 0.35  division  display amplitude).
e.  CHECK—that  the  display  cannot  be  triggered  In  either  HF  REJ  of 
NOISE  REJ  CPLG.
f.  Set:
Horizontal  MODE  B
A/B  SELECT  B Trigger
B  SEC/DIV  1  |1S
g.  CHECK—that the display cannot be triggered in HF  REJ or NOISE REJ 
CPLG by  adjusting the Trigger  LEVEL control.
3.  25  MHz Trigger  Sensitivity
a.  Set:
A
A Trigger 
DC 
50 ns
Horizontal  MODE 
A/B  SELECT 
TRIGGER CPLG 
A SEC/DIV
b.  Remove the  10X  and  2X  BNC attenuators from the  signal path.
c.  Set  leveled sine-wave  generator  output  to  produce  a 7.0  division  dis
play amplitude  at  25  MHz.
d.  Add a 10X and a 2X BNC attenuator before the 50 f l   BNC termination.
e.  CHECK—that  the  display  Is  stably  triggered  In  DC,  LF  REJ,  and  AC 
Trigger CPLG;  the  display  is  not  triggered  in  NOISE  REJ  and  HF  REJ 
Trigger CPLG settings.
f.  Set:
TRIGGER CPLG 
Horizontal  MODE 
A/B SELECT 
B SEC/DIV
DC
B
B Trigger 
20 ns
7-30 2245A  Operators