Performance  Check  Procedure
TRIGGERING
Equipment  Required  (See Table 7-1) 
Leveled sine-wave  generator 
50  11  BNC  coaxial cable 
2X  BNC  attenuator 
Dual-input  coupler
Function generator 
10X  BNC attenuator 
50  Cl  BNC termination 
TV signal generator
1.  500  Hz Trigger  Sensitivity
a.  Set:
READOUT  (Intensity) For a viewable  readout
A  INTEN
For a viewable trace
VERTICAL  MODE CH  1
CH  1  and CH  2  Input  COUPLING
DC
CH  1  VOLTS/DIV 0.1  V
SCOPE  BW On
Horizontal MODE
A
A SEC/DIV
1  ms
Horizontal  POSITION
12  o'clock
A/B SELECT A Trigger
TRIGGER MODE
AUTO  LEVEL
TRIGGER SOURCE
VERT
TRIGGER CPLG
AC
TRIGGER SLOPE
(positive-going)
TRIGGER HOLDOFF
Min
FOCUS
For best defined  display
CLEAR  MEAS'MT
Press to remove  measurement 
cursors.
b.  Connect  function  generator  (FG  502)  output  to  the  CH  1  input  via  a 
50 f l  BNC coaxial  cable and  a  50 Cl  BNC  termination.
c.  Set function generator  (FG 502)  output to produce a 7.0 division sine- 
wave  display at  500  Hz.
d.  Add  a  10X and a 2X  BNC attenuator before the 50 Cl  BNC termination 
(for a 0.35  division display).
7-28
2245A  Operators