Performance  Check  Procedure
c.  Connect calibration generator  (PG 506)  fast rise,  rising-edge signal to 
the  CH  1  and  CH  2  inputs via a 50  O  precision  BNC coaxial  cable,  a 
50 f l   BNC termination,  and  a BNC dual-input  coupler.
d.  Connect  calibration  generator  trig  out  signal  to  the  CH  3  input  via  a 
50 f l   BNC coaxial  cable and  a 50 Cl  BNC termination.
e.  Set  the  calibration  generator to  Fast  Rise  and adjust  Pulse Amplitude 
for five divisions of signal  amplitude  at  1  MHz.
f.  Position the rising edges of the superimposed waveforms horizontally to 
the center vertical  graticule line.
NOTE
It  may  be  necessary  to  readjust  the  trigger  level  to  display  the 
rising-edge signal In the Delay Match steps.  This can be done by 
pressing  the  upper  Trigger  MODE  button  to  cause  the  trigger 
level  to be automatically readjusted.
g.  Set  XI0  MAG  On  (for 2  ns/div sweep speed).
h.  CHECK—that  the  leading  edges  of the  two waveforms have less than 
0.1  horizontal  division  separation  at  the  center graticule  line  excluding 
trace width.
14.  CH  1  to  CH  4  Signal  Delay  Match
a.  Set VERTICAL  MODE to  CH  1  and CH  4  (CH  2  off).
b.  Move the  CH  2  signal to the CH 4 Input connector.
c.  Superimpose the  CH 4 waveform on the  CH  1  waveform.
d.  CHECK—that  the  leading edges  of the  two  waveforms have  less than 
0.1  horizontal  division separation  at  the  center graticule  line  excluding 
trace width.
15.  CH  3  to  CH 4  Signal  Delay  Match
a.  Set:
VERTICAL MODE  CH  3 and CH 4  (CH  1  off)
TRIGGER SOURCE  CH 2
b.  Move the CH  3 signal to the  CH  2  input and the CH  1  trigger signal to 
the  CH 3 input.
7-20
2245A  Operators