Performance  Check Procedure
TRIGGER CPLG  DC
g.  Move test signal from CH  1  to the  CH 2  input.
h.  Set  leveled  sine-wave  generator  output  to  produce  a  1.0  division  dis
play  amplitude  at  150  MHz.
I.  CHECK—that  a  stable  display  can  be  obtained.  (The  Trigger  LEVEL
control may  be adjusted to  improve the  display stability.)
j.  Repeat  the  procedure  for  CH  3  and  CH  4  (turn  on  the  appropriate 
VERTICAL  MODE  and move the test  signal  as required).
k.  Move test  signal to the  CH  1  input.
l.  Set VERTICAL  MODE to CH  1  (others  off).
m.  Remove the  2X  BNC attenuator from the test signal  path.
n.  Set leveled sine-wave generator output for a 2.2 division display ampli
tude  at  100  MHz.
o.  CHECK—that  the  display  is  stably  triggered  with  NOISE  REJ  Trigger 
CPLG  but is  not  triggered with  HF REJ  Trigger CPLG.
p.  Set leveled sine-wave generator output for a 0.5 division display ampli
tude  at  100  MHz.
q.  CHECK—that the display is not triggered in NOISE REJ Trigger CPLG.
r.  Set:
TRIGGER CPLG  DC
Horizontal MODE  B
A/B SELECT  B Trigger
s.  Repeat  100  MHz  NOISE  REJ  Trigger  CPLG  procedure  for  the  B 
Trigger.
5.  Single  Sweep  Mode
a.  Set:
Horizontal  MODE 
A SEC/DIV 
A/B  SELECT
A
10 (is 
A Trigger
7-32
2245A  Operators