Performance  Check Procedure
OPTION  15
Equipment  Required  (See  Table 7-1)
Calibration  generator  Two  50  f l  BNC coaxial cables
Test oscilloscope  50  f l  Precision  BNC coaxial  cable
Level sine-wave  generator  Two  50  f l   BNC termination
1.  Signal  Output
a.  Set:
VERTICAL  MODE 
CH  1  and  CH 2 
CH  3  and CH 4 
BW  LIMIT
VOLTS/DIV 
CH  1  and  CH 2
Input Coupling 
CH  1  and CH  2
A and B SEC/DIV
TRIGGER  MODE
SOURCE
COUPLING
On  (light  on) 
Off  (light off) 
Off  (light  off)
2  MV
GND 
1  ms
AUTO  LVL 
VERT 
NOISE  REJ
b.  Push the  CH  2  VERTICAL MODE button so that  light  is  off.
c.  Connect the CH 2  signal from the  rear-panel CH 2 SIGNAL OUT con
nector to the  CH  1  OR X input connector via a 50 f l  BNC cable.
d.  Align the  CH  1  trace to the center graticule line.
e.  Set  CH  1  Input Coupling to  DC.
f.  CHECK—Display  amplitude  to  4.5  to  5.5  divisions  (neglect  trace 
width).
g.  Connect  a  1  kHz,  10  mV  standard-amplitude  signal  from  the  Calibra
tion  Generator to  the CH  2  Input  Connector via a 50 f l BNC cable.
7-50 2245A  Operators