Performance  Check  Procedure
MEASUREMENT CURSORS
Equipment  Required  (See  Table 7-1)
Time mark generator
Calibration generator
50 f l   BNC coaxial  cable
50  f l  BNC termination
1. |<- SEC -»l and l«- 1/SEC -H  Cursor Accuracy
Set:
READOUT  (Intensity)
For a viewable  readout
A INTEN
For a viewable trace
VERTICAL  MODE
CH  1
CH  1  VOLTS/DIV 0.5 V
CH  1  and CH  2
Input  COUPLING
DC
CH  2  INVERT
Off
SCOPE  BW
Off
Horizontal MODE
A
A SEC/DIV
1  ms
A/B  SELECT
A Trigger
TRIGGER MODE
AUTO  LEVEL
TRIGGER  CPLG
DC
TRIGGER SOURCE
VERT
TRIGGER SLOPE
(positive-going)
TRIGGER HOLDOFF
Min
FOCUS
For best  defined  display
b.  Connect time-mark generator  (TG 501)  output via a 50 f l  BNC coaxial 
cable  and  a 50 f l   BNC termination to the  CH  1  input.
c.  Set time-mark generator for  1  ms time marks.
d.  Position first  time  marker horizontally  to the  first vertical  graticule line 
(left-most  edge of the  graticule).
e.  Press TIME MEASUREMENTS button to display the H- SEC -»l cursors.
f.  Position the reference cursor to the second time marker and the delta 
cursor to the tenth time  marker.
7-46
2245A  Operators