EasyManua.ls Logo

Tektronix 2245A - MEASUREMENT CURSORS; k- SEC ~>l and l<- 1;SEC -| Cursor Accuracy

Tektronix 2245A
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Check Procedure
MEASUREMENT CURSORS
Equipment Required (See Table 7-1)
Time mark generator
Calibration generator
50 f l BNC coaxial cable
50 f l BNC termination
1. |<- SEC -»l and l«- 1/SEC -H Cursor Accuracy
Set:
READOUT (Intensity)
For a viewable readout
A INTEN
For a viewable trace
VERTICAL MODE
CH 1
CH 1 VOLTS/DIV 0.5 V
CH 1 and CH 2
Input COUPLING
DC
CH 2 INVERT
Off
SCOPE BW
Off
Horizontal MODE
A
A SEC/DIV
1 ms
A/B SELECT
A Trigger
TRIGGER MODE
AUTO LEVEL
TRIGGER CPLG
DC
TRIGGER SOURCE
VERT
TRIGGER SLOPE
(positive-going)
TRIGGER HOLDOFF
Min
FOCUS
For best defined display
b. Connect time-mark generator (TG 501) output via a 50 f l BNC coaxial
cable and a 50 f l BNC termination to the CH 1 input.
c. Set time-mark generator for 1 ms time marks.
d. Position first time marker horizontally to the first vertical graticule line
(left-most edge of the graticule).
e. Press TIME MEASUREMENTS button to display the H- SEC -»l cursors.
f. Position the reference cursor to the second time marker and the delta
cursor to the tenth time marker.
7-46
2245A Operators

Table of Contents

Related product manuals