Performance  Check Procedure
e.  Set A  INTEN  to midrange.
f.  Disconnect the  test equipment  from the  2245A.
2.  PROBE  ADJUST  O utput
a.  Set:
CH  1  VOLTS/DIV  10  mV
SEC/DIV  0.2  ms
b.  Connect  a  10X  probe  to  the  CH  1  input  connector  and  connect  the 
probe tip to the 2245A PROBE ADJUST output.  (When using Tektronix 
coded probes,  the  readout  changes to  .1  V.)
c.  CHECK—For  a  5-division  vertical  display  of  PROBE  ADJUST  square- 
wave  signal  (square-wave  period is  typically  1  ms,  within 25%).
3.  AUTO  SETUP  Functional  Check
a.  Set:
CH  1  COUPLINQ  GND
CH  1  VOLTS/DIV  2  mV
A  SEC/DIV  20 ns
b.  Press the  AUTO  SETUP button.
c.  Check that the Probe Adjust waveform is stably displayed on the upper 
half of the  crt.