Performance  Check  Procedure
c.  Superimpose CH 3 and CH 4 waveforms at the center graticule line.
d.  CHECK—that  the  leading  edges  of the  two waveforms have  less than
0.1  horizontal division separation at the  center graticule line.
e.  Disconnect the test  setup.
16.  CH  1  and  CH 2 Vertical Bandwidth
a.  Set:
X10  MAG 
VERTICAL  MODE 
SEC/DIV
CH  1  VOLTS/DIV 
CH  1  and CH  2  Input COUPLING 
TRIGGER SOURCE 
Horizontal POSITION
Off
CH  1  (CH  3 and CH 4  off)
0.1  ms
5  mV
DC
VERT
12  o’clock
b.  Connect  leveled sine-wave  generator  (SG  503)  output to the CH  1  in
put via a 50 f l   precision  BNC coaxial cable and a 50 f l   BNC termina
tion.
c.  Set  the  leveled  sine-wave  Generator  output  for  a  six-division  signal 
amplitude  at  50 kHz.
d.  Set  the  generator  Frequency  Range  and  Frequency  Variable  controls 
for a 90  MHz  output  signal.
e.  CHECK—the displayed signal amplitude  Is 4.2  divisions or more  as the 
frequency  is  Increased to  100  MHz.
f.  Repeat  the  frequency  setup  and  CHECK  procedure  for  VOLTS/DIV 
settings  of 50 mV through  0.5 V.
g.  Move the test  signal to the  CH  2  input.
h.  Set:
VERTICAL  MODE  CH  2  (CH  1  off)
CH  2  VOLTS/DIV  5  mV
I.  Repeat the  complete  Bandwidth check procedure for Channel 2.
2245A  Operators
7-21