1. Push the Utility button.
2. Push the Calibration & Diagnostic side-menu button.
3. Turn the Multipurpose knob to choose Diagnostics, then click the
Multipurpose knob to select.
4. Push the Self Test side-menu button to display the Self Test menu and Self
Test Results readout.
5. Turn the Multipurpose knob to select Run Self Test, then click the
Multipurpose knob to start testing using the default settings. the self test
takes a few seconds to run.
6. Use the other Self Test menu items to refine how to run the self test; you can
run the tests a specified number of times (Loop Times), run tests forever
(Loop Forever), run tests until a test fails and stop testing (Loop Until Fail),
or run a test on a failed tests (Loop On Failure).
NOTE. The oscilloscope resets to the factory default settings after running the self
tests.