Application Examples Typical Equipment Setup-Resume Measurement
TDSUSBF compliance test xture (Revision B)
Two of the following single-ended voltage probes: P6245, P6243, TDP1500, TDP3500, or TAP1500.
Typical Equipment Setup-Resume Measurement
To set up the Device SQ in the test xture for the Resume test, follow these steps:
1. Set the S6 switch to the Init position.
2. Use a standard length of the USB cable with an A plug on one end and a B plug on the other end.
Connect one end of the cable to the B receptacle socket connector on the Init port of the Device SQ
section and the other end to the host port A socket.
3. Connect the A receptacle from the Device SQ test port to the unit under test (device).
4. Connect the single-ended probes of the oscilloscope to the D+ and D– pins.
5. Select the Resume measurement in the application, congure its options and select the Run button
to run the application.
6. Select OK after acquiring a waveform. Verify that it is a correct waveform.
TDSUSB2 Universal Serial Bus Measurements Package 97