Operating Basics Conguring Suspend Measurement
3. Set the input Signal Direction to either Upstream or Downstream.
4. Select th
e channel source in the D+ and D– elds. Use the drop down arrow in the D+, D– eldtoset
the source. The available options are: Ch1- Ch4 and Ref1-Ref4.
5. Select t
he Run button.
6. To generate reports, select Utilities > Report Generator.
See Also
Gener
ating Reports
Conguring Suspend Measurement
To congure Suspend measurement, follow these steps:
1. Select the High Speed measurement tab > More button > Suspend measurement. This measures the
time between the end of the last SOF and the rising edge transition to the Full Speed J state. The
ac
ceptable range is between 3 ms to 3.125 ms.
2. To access the Congure Measurements menu, go to Measurements > Congure.
42 TDSUSB2 Universal Serial Bus Measurements Package