Reference Reset from High Speed Test Equipment Setup
6. Select the Suspend measurement, congure its options and select the Run button to run the application.
7. Select OK after acquiring a waveform. Verify that it is a correct waveform.
Reset
from High Speed Test Equipment Setup
To se
t up the Device SQ in the test xture for the Reset from High Speed test, follow these steps:
1. Set the S6 switch to the Init position.
2. Use a standard length of the USB cable with an A plug on one end and a B plug on the other end.
3. Connect one end of the cable to the B socket on the Init port of the Device SQ section and the other
end to the host port A socket.
4. Connect the A receptacle from the Device SQ test port to the unit under test (device).
5. Connect the single-ended probes of the oscilloscope to the D+ and D– pins.