Operating Instructions Model 200E Instruction Manual
To disable a data channel, press SETUP-DAS-ENTR and the following keys:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
CHANNEL ENABLE:ON
<SET SET>
EDIT
PRINT EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CHANNEL ENABLE:ON
OFF
ENTR EXIT
Toggle key to turn
channel ON or OFF
Press
SET>
key until…
6.10.2.12. HOLDOFF Feature
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during
the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.5). To enable or
disable the HOLDOFF for any one iDAS channel, press SETUP-DAS-ENTR and then:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
CAL HOLD OFF:ON
SET>
EDIT
PRINT EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CAL HOLD OFF:ON
ON
ENTR EXIT
Toggle key to turn
HOLDOFF ON or OFF
Press
SET>
key until…
100 044100102 Rev A